Please use this identifier to cite or link to this item: http://hdl.handle.net/10773/22205
Full metadata record
DC FieldValueLanguage
dc.contributor.authorCruz, Joanapt
dc.contributor.authorBrooks, Dinapt
dc.contributor.authorMarques, Aldapt
dc.date.accessioned2018-02-15T15:17:25Z-
dc.date.available2018-02-15T15:17:25Z-
dc.date.issued2017-04-
dc.identifier.issn0022-4707pt
dc.identifier.urihttp://hdl.handle.net/10773/22205-
dc.description.abstractThis study aimed to: assess step-count accuracy of a piezoeletric pedometer (Yamax PW/EX-510), when worn at different body parts, and a triaxial accelerometer (GT3X+); compare device accuracy; and identify the preferred location(s) to wear a pedometer. METHODS: Sixty-three healthy adults (45.8±20.6 years old) wore 7 pedometers (neck, lateral right and left of the waist, front right and left of the waist, front pockets of the trousers) and 1 accelerometer (over the right hip), while walking 120m at slow, self-preferred/normal and fast paces. Steps were recorded. Participants identified their preferred location(s) to wear the pedometer. Absolute percent error (APE) and Bland and Altman (BA) method were used to assess device accuracy (criterion measure: manual counts) and BA method for device comparisons. RESULTS: Pedometer APE was below 3% at normal and fast paces despite wearing location, but higher at slow pace (4.5–9.1%). Pedometers were more accurate at the front waist and inside the pockets. Accelerometer APE was higher than pedometer APE (P<0.05); nevertheless, limits of agreement between devices were relatively small. Preferred wearing locations were inside the front right (n=25) and left (n=20) pockets of the trousers. CONCLUSION: Yamax PW/EX-510 pedometers may be preferable than GT3X+ accelerometers to count steps, as they provide more accurate results. These pedometers should be worn at the front right or left positions of the waist or inside the front pockets of the trousers.pt
dc.language.isoengpt
dc.publisherEdizioni Minerva Medicapt
dc.relationinfo:eu-repo/grantAgreement/FCT/SFRH/SFRH%2FBD%2F81328%2F2011/PTpt
dc.rightsopenAccesspor
dc.subjectActigraphypt
dc.subjectDimensional measurement accuracypt
dc.subjectExercisept
dc.subjectWalkingpt
dc.titleAccuracy of piezoelectric pedometer and accelerometer step countspt
dc.typearticlept
dc.peerreviewedyespt
ua.distributioninternationalpt
degois.publication.firstPage426pt
degois.publication.issue4pt
degois.publication.lastPage433pt
degois.publication.titleThe Journal of Sports Medicine and Physical Fitnesspt
degois.publication.volume57pt
dc.identifier.doi10.23736/S0022-4707.16.06177-Xpt
Appears in Collections:ESSUA - Artigos
Lab3R - Artigos

Files in This Item:
File Description SizeFormat 
Cruz2016_Accuracy of piezoelectric pedometers and accelerometer.pdf437.43 kBAdobe PDFView/Open


FacebookTwitterLinkedIn
Formato BibTex MendeleyEndnote Degois 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.