Please use this identifier to cite or link to this item: http://hdl.handle.net/10773/5261
Title: Robustness of the co-ion transfer ratio in capillary electrophoresis
Author: Erny, Guillaume L.
Esteves, Valdemar I.
Keywords: Indirect detection
Organic acid
Repeatability
Robustness
Transfer ratio
Issue Date: 2009
Publisher: Wiley-VCH Verlag Berlin
Abstract: In CE, indirect detection mode often exhibits a lower precision than its direct counterpart. Although various explanations have already been advanced, in this work, we aimed to investigate if this is due, in part, to problems of robustness of the coion transfer ratio (TR), thus being inherent to this particular detection scheme. This was investigated using simulation software that allows an accurate control of various parameters and validated using acetic acid as a test compound. It was conclusively demonstrated that the TR could vary by more than 6% when the concentration of one of the ions in the BGE was changed by as few as 1%. The presence of a system peak seems to be particularly damaging as it has been shown that the TR of peaks whose mobilities differ by more than 0.5610– 8 m2V– 1 s – 1 from one of the system peaks, still have a relatively low robustness.
Peer review: yes
URI: http://hdl.handle.net/10773/5261
DOI: 10.1002/jssc.200900069
ISSN: 1615-9306
Appears in Collections:DQ - Artigos

Files in This Item:
File Description SizeFormat 
JSS_Set_2009.pdf512.01 kBAdobe PDFrestrictedAccess


FacebookTwitterLinkedIn
Formato BibTex MendeleyEndnote Degois 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.