dc.description.abstractDetermining the usability of mechanical ventilator interfaces is vital in ensuring correct usage under critical procedures that can place patients’ health in danger. Heuristic evaluation has been used to evaluate usability in many different areas, including medical devices. The objective of this study is to evaluate the usability of a mechanical ventilator using the heuristic evaluation method. The evaluation of a mechanical ventilator simulator (Drager's Evita V500) was performed using the Heuristic Evaluation System Checklist (HESC). Three evaluators took part in the experiment in assessing the usability through HESC. The results show that of the 292 checklist sub-heuristics included in the HESC Checklist, 127 were compliant (had no flaws), 75 were not compliant (had flaws) and 95 were not applicable. The results point out that the mechanical ventilator interfaces should be improved to avoid human error due to usability issues. The results also show that heuristic evaluation generates a large quantity of objective information, which provides an exhaustive identification of aspects that should be improved.pt_PT
dc.language.isoengpt_PT
dc.publisherElsevierpt_PT
dc.rightsopenAccesspt_PT
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/pt_PT
dc.subjectUsabilitypt_PT
dc.subjectVentilatorpt_PT
dc.subjectHeuristic evaluationpt_PT
dc.subjectMedical devicespt_PT
dc.subjectHuman-Computer Interactionpt_PT
dc.titleHeuristic evaluation of the usability of a mechanical ventilator Interface through a simulatorpt_PT
dc.typearticlept_PT
dc.description.versionpublishedpt_PT
dc.peerreviewedyespt_PT
degois.publication.firstPage1232pt_PT
degois.publication.lastPage1239pt_PT
degois.publication.titleProcedia Computer Sciencept_PT
degois.publication.volume219pt_PT
dc.identifier.doi10.1016/j.procs.2023.01.406pt_PT
Appears in Collections:IEETA - Artigos
DCM - Artigos
CINTESIS - Artigos

Files in This Item:
File Description SizeFormat 
Heuristic-Evaluation-of-the-Usability-of-a-Mechanical-Ve_2023_Procedia-Compu.pdf560.81 kBAdobe PDFView/Open


FacebookTwitterLinkedIn
Formato BibTex MendeleyEndnote Degois 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.