Please use this identifier to cite or link to this item: http://hdl.handle.net/10773/4363
Full metadata record
DC FieldValueLanguage
dc.contributor.advisorVaz, Francisco António Cardosopt
dc.contributor.advisorFish, Peterpt
dc.contributor.authorBastos, Carlos Alberto da Costapt
dc.date.accessioned2011-11-16T11:41:06Z-
dc.date.available2011-11-16T11:41:06Z-
dc.date.issued1999-
dc.identifier.urihttp://hdl.handle.net/10773/4363-
dc.descriptionDoutoramento em Engenharia Electrotécnicapt
dc.description.abstractO detector ultra-sónico de fluxo sanguíneo usa o efeito Doppler para estimar de forma não invasiva a velocidade do sangue na circulação. Tem sido bastante usado nas ultimas quatro décadas para detectar a presença de estenoses. O desenvolvimento de novas técnicas de processamento do sinal Doppler necessita de sinais de teste cujas características sejam conhecidas ou possam ser medidas com precisão. Isto é difícil de obter com sinais Doppler medidos in vivo devido à elevada variação do fluxo sanguíneo de pessoa para pessoa e também com o estado fisiológico da pessoa no momento da medida, por exemplo a tensão arterial influencia significativamente o fluxo sanguíneo. Um modelo para gerar sinais Doppler simulados cujas características sejam controláveis e/ou mensuráveis é uma ferramenta bastante útil, pois permite que as novas técnicas de processamento do sinal Doppler sejam testadas em condições controladas. Permite, também, estudar o efeito de viários factores que afectam o espectro do sinal Doppler. Habitualmente o efeito individual dos viários factores não pode ser identificado quando são usados sinais medidos in vivo. Neste trabalho foi desenvolvido um modelo para gerar sinais Doppler ultra-sónicos simulados. O modelo contêm dois sub-modelos, um para o fluxo sanguíneo nos membros inferiores de um ser humano e outro para gerar os sinais simulados a partir do campo de velocidades do sangue e das características do instrumento. O fluxo sanguíneo nos membros inferiores foi simulado com um análogo eléctrico para a rede vascular dos membros inferiores. Cada artéria foi simulada por uma linha de transmissão com perdas e as redes vasculares periféricas por um circuito Windkessel com três elementos. O circuito eléctrico foi implementado com o simulador de circuitos SPICE. Para simular a interacção entre os glóbulos vermelhos e o campo de ultra-sons o vaso sanguíneo foi dividido em pequenos volumes elementares. As contribuições dos volumes elementares foram todas somadas para gerar o sinal Doppler simulado. O modelo fez algumas aproximações como sejam, por exemplo, considerar o fluxo sanguíneo laminar e sem rotação. As características dos sinais gerados pelo modelo são bastante parecidas com as esperadas para o sinal Doppler real. O modelo desenvolvido foi usado para estudar a influencia que a aceleração sanguínea, o tamanho do volume de amostragem e a duração da janela de amostragem têm na largura de banda eficaz do espectro do sinal Doppler. Foi deduzida uma formula que estima a largura de banda eficaz a partir das contribuições individuais do alargamento espectral devido à não estacionaridade, do alargamento espectral intrínseco, do alargamento espectral devido à duração da janela de amostragem e ainda da gama das velocidades que passam pelo volume de amostragem. Foram, ainda, deduzidas expressões em forma fechada para o espectro de potência do sinal Doppler devido unicamente à gama de velocidades que atravessam um volume de amostragem com forma Gaussiana colocado num perfil de velocidades com forma exponencial. Foram, também, obtidas expressões para a largura de banda eficaz no caso especial do volume de amostragem Gaussiano ter simetria esférica e estar colocado no centro do vaso sanguíneo.pt
dc.description.abstractThe Doppler ultrasonic blood ow detector estimates non-invasively the velocity of blood in the circulatory system. It has been extensively used in the last four decades for the detection of stenoses in the circulation. The development of new signal processing techniques for the Doppler signal requires test signals with known or measurable characteristics. This is very di cult to achieve with Doppler signals obtained in vivo because of the variability of blood ow between persons and with physiological state, for example blood pressure. A model for generating simulated Doppler signals whose characteristics are controllable and/or measurable is a useful tool because it permits the test of new processing techniques under controlled conditions. It permits also the study of the e ect of various factors on the Doppler spectrum. Usually these e ects cannot be isolated with in vivo measurements. During this work a model for the generation of simulated Doppler ultrasound signals was developed. It comprised two sub-models one for blood ow in the human lower limb and the other for generating simulated signals from the blood velocity eld and the instrument's characteristics. Blood ow in the lower limb was modelled by an electric analogue for the lower limb vascular tree. Each artery was modelled by a lossy transmission line and the peripheral vascular beds by three{element Windkessel models. The electric analogue circuit was implemented with the SPICE circuit simulator. To simulate the inter-action of the blood cells with the ultrasonic eld the vessel was divided into small elemental volumes whose contributions were added together to generate the simulated Doppler signal. The model assumed irrotational laminar ow and some other simplifying approximations. The characteristics of the signals generated by the model were similar to those expected for the Doppler signal. The model was used to study the in- uence of blood acceleration, sample volume size and data segment duration on the root mean square (rms) width of the Doppler spectrum. A simple formula was derived for estimating the Doppler rms spectral width from the individual contribution of non-stationarity broadening, intrinsic broadening, window broadening and the range of blood velocities passing through the sample volume. In addition closed form expressions were derived for the Doppler power spectrum due solely to the range of blood velocities passing through a Gaussian sample volumes placed in irrotational laminar ow with a velocity pro le obeying a simple power law. Closed form expressions were also obtained for the root mean square spectral width in the special case of a spherically symmetric Gaussian sample volume placed in the centre of the vessel.pt
dc.language.isoengpt
dc.publisherUniversidade de Aveiropt
dc.rightsopenAccesspor
dc.subjectEngenharia electrotécnicapt
dc.subjectFluxo sanguíneo - Monitorizaçãopt
dc.subjectUltrassonografia de Doppler - Aplicações em medicinapt
dc.subjectEngenharia biomédicapt
dc.titleA model for the simulation of Doppler ultrasound signal from pulsatic blood flowpt
dc.title.alternativeUm modelo para a simulação de sinais Doppler ultra-sonicos provenientes de fluxo sanguínio pulsátilpt
dc.typedoctoralThesispt
thesis.degree.leveldoutoramentopt
thesis.degree.grantorUniversidade de Aveiropt
dc.identifier.tid101077165-
Appears in Collections:UA - Teses de doutoramento
DETI - Teses de doutoramento

Files in This Item:
File Description SizeFormat 
100.pdf1.9 MBAdobe PDFView/Open


FacebookTwitterLinkedIn
Formato BibTex MendeleyEndnote Degois 

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.