Título: Comparação dosimétrica entre MLC e MICRO-MLC: estudo de casos de radiocirurgia (SRS)
Autor: Leitão, Sílvia dos Anjos Pinheiro
Orientador: Silva, Augusto Marques Ferreira da
Silva, Joana Vale Ferreira da
Faria, David
Palavras-chave: Radioterapia
Radiocirurgia
Metástases
Cérebro
Meningiomas
Neurinomas
Colimador multilâminas
Micro Colimador
Apex™
AgilityTM
Data de Defesa: 13-Jan-2021
Resumo: Enquadramento: O objetivo dos tratamentos de radioterapia é irradiar o volume alvo minimizando a dose recebida pelos órgãos de risco adjacentes (OAR’s). Uma correta colimação do feixe de radiação ao volume alvo utilizando o colimador multilâminas (MLC) permite reduzir a dose nos OAR’s para valores abaixo dos valores de tolerância impostos pela radiobiologia tecidular. O MLC consiste em vários pares de folhas que se movem de forma independente e que permitem bloquear a radiação do feixe incidente. Diferentes MLCs, com diferentes características, permitem executar planos de tratamento e administração de doses de forma diferente. Objetivos: Comparar as características de dois colimadores multilâminas e avaliar as vantagens dosimétricas da utilização de cada um, no planeamento de tratamentos de Radiocirurgia, consoante a patologia, forma e volume do tumor. Métodos: Foram estudados 30 casos de radiocirurgia cujo plano foi executado com o MLC Agility™ no sistema de planeamento (TPS) Monaco®, utilizando o algoritmo Monte Carlo para o cálculo de planimetrias com a técnica de Volumetric Modulated Arc Therapy (VMAT). Utilizando planos já executados com MLC Agility™ foi feita a sua reotimização, mantendo a configuração de campos, as mesmas funções objetivo de otimização, mas com o MLC Apex™. Resultados: Os casos clínicos foram tratados com uma dose total entre 12 e 20 Gy numa única fração e 80% (n=24) das lesões eram irregulares com um volume tumoral que variava entre 0,06 e 15,21 cm3 (mediana= 2,68 cm3 ). Para os 30 planos de SRS VMAT, ambos os colimadores apresentaram resultados semelhantes em termos de cobertura do PTV clinicamente aceitável para o tratamento de SRS (p= 0.84, n= 60). Não se observaram diferenças significativas nos índices da qualidade, exceto no índice de gradiente, onde se observa uma melhoria estatisticamente significativa com o micro-MLC ApexTM, mostrando alguma vantagem clínica nos neurinomas e alguns meningiomas. Conclusão: O tratamento de SRS com o micro-MLC Apex™ apresentou resultados dosimétricos muito semelhantes ao MLC Agility™, em termos de cobertura do PTV, índices da qualidade do plano e limites de dose a cumprir nos OAR’s. Não houve diferenças estatisticamente significativas nos índices da qualidade, exceto no índice de gradiente, em que se observa uma melhoria com o micro-MLC Apex™, que nos permite concluir que terá uma vantagem clínica mais significativa nos neurinomas devido ao menor tamanho desses tumores.
Background: The goal of radiotherapy treatments is to irradiate the target volume minimizing the dose received by the adjacent organs at risk (OAR’s). Correct beam collimation to the target volume using the multi-leaf collimator (MLC) reduces the dose in OAR’s below the tolerance values imposed by tissue radiobiology. The MLC consists of several independently moving leaf pairs that block the incident radiation beam. Different MLCs, with distinct characteristics, allow the execution of treatment plans and administration of different doses. Purpose: Comparison of the characteristics of two multi-leaf collimators and evaluation of the dosimetric advantages of their use in the planning of radiosurgery treatments, according to the pathology, shape and volume of the tumour. Methods: Thirty cases of radiosurgery whose planning was performed with MLC Agility™ in the Monaco® planning system (TPS), using the Monte Carlo algorithm for the calculation of planimetry using the technique Volumetric Modulated Arc Therapy (VMAT), were studied. Plans will be reoptimized, maintaining field configuration, and the same objectives and optimization functions, but with MLC Apex™. Results: The clinical cases were treated with a prescription dose of 12 to 20 Gy in a single fraction and 80% (n=24) of the lesions included in the initial review were irregular, with gross tumour volume ranging from 0.06 to 15.21 cm3 (median: 2.68 cm3 ). For the 30 VMAT SRS plans, both collimators showed comparable results in terms of clinical acceptable PTV coverage objective for SRS treatment (p= 0.84, n=60). There was no significant difference in the quality metrics except in the gradient index, where a statistically significant improvement is observed with the micro-MLC ApexTM, showing some clinical advantage in neurinomas and some meningiomas. Conclusion: The treatment of the SRS with the micro-MLC Apex™ shows results very similar to the MLC Agility™, in terms of PTV coverage, plan quality metrics and dose limits the sparing of adjacent organs. There were no statistically significant differences in the quality metrics, except in the gradient index, in which an improvement is observed with the micro-MLC Apex™, which allows us to conclude that it will have a more significant clinical advantage in neurinomas due to the smaller size of these tumors.
URI: http://hdl.handle.net/10773/30562
Aparece nas coleções: DCM - Dissertações de mestrado
UA - Dissertações de mestrado
DETI - Dissertações de mestrado
DFis - Dissertações de mestrado
ESSUA - Dissertações de mestrado
DEGEIT - Dissertações de mestrado

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
Documento_Sílvia_Leitão.pdf1.55 MBAdobe PDFVer/Abrir


FacebookTwitterLinkedIn
Formato BibTex MendeleyEndnote Degois 

Todos os registos no repositório estão protegidos por leis de copyright, com todos os direitos reservados.